Repeated readings vary unpredictably around a stable value. Which error is illustrated?
A Gross error
B Random error
C Zero error
D Calibration offset
Answer: B. Random error. Random errors cause scatter and can be reduced statistically by repeated measurements.
A voltmeter always reads 2 V too high across its range. What kind of error is this?
A Random error
B Parallax only
C Quantization noise
D Systematic error
Answer: D. Systematic error. A persistent bias is systematic and calls for calibration or correction.
A mirror scale on an analog meter helps reduce which reading error?
A Frequency error
B Burden error
C Hysteresis loss
D Parallax error
Answer: D. Parallax error. Aligning the pointer with its reflection keeps the eye normal to the scale.
A voltmeter is connected across a high-resistance circuit and changes the voltage. What is this called?
A Creeping
B Phase reversal
C Loading effect
D Mutual induction
Answer: C. Loading effect. Finite input resistance draws current and alters the circuit being measured.
A 10 kΩ voltmeter is placed across a 10 kΩ resistor. What is their equivalent resistance?
A 10 kΩ
B 5 kΩ
C 20 kΩ
D 100 kΩ
Answer: B. 5 kΩ. Two equal 10 kΩ resistances in parallel give 5 kΩ.
For minimum circuit disturbance, what should a voltmeter input resistance be?
A Very low
B Exactly zero
C Equal to source resistance
D Very high
Answer: D. Very high. A high input resistance reduces the current drawn from the measured circuit.
For minimum series voltage drop, an ammeter should have which resistance?
A Very high
B Very low
C Infinite
D Equal to load resistance
Answer: B. Very low. An ammeter is inserted in series, so its resistance should be small.
What must be avoided with an energized current transformer primary?
A Open CT secondary
B Short CT secondary
C Connect rated burden
D Use a shorting link
Answer: A. Open CT secondary. Opening the CT secondary can induce a dangerously high voltage.
Which wattmeter coil is placed in series with the load?
A Pressure coil
B Potential coil
C Current coil
D Voltage coil
Answer: C. Current coil. The current coil carries load current and is connected in series.
Which wattmeter coil is connected across the load?
A Potential coil
B Current coil
C Shunt of ammeter
D CT primary
Answer: A. Potential coil. The potential or pressure coil senses load voltage in parallel.
In the two-wattmeter method, how is total three-phase active power found?
A Subtract both readings
B Multiply both readings
C Divide by √3
D Add both readings
Answer: D. Add both readings. The algebraic sum of the two readings gives total active power for a three-wire system.
For a balanced three-phase load at unity power factor, how do the two wattmeter readings compare?
A They are equal
B One is zero
C They are opposite
D Both are negative
Answer: A. They are equal. At φ=0, the two wattmeters measure equal positive power.
At what balanced-load power factor does one wattmeter read zero in the two-wattmeter method?
A 1.0
B 0.866
C 0.5
D 0
Answer: C. 0.5. At φ=60°, cos(30°+φ)=0, so one reading is zero and cosφ=0.5.
At a balanced-load power factor below 0.5, what happens to one wattmeter reading?
A Both become zero
B Both stay equal
C It becomes negative
D It doubles exactly
Answer: C. It becomes negative. For φ greater than 60°, one wattmeter has a negative algebraic reading.
An energy meter registers with no connected load. What is this fault called?
A Loading
B Damping
C Aliasing
D Creeping
Answer: D. Creeping. Creeping is slow registration of an energy meter at no load.
What is the main purpose of calibrating a measuring instrument?
A Increase supply voltage
B Remove its scale
C Compare with reference
D Change its unit
Answer: C. Compare with reference. Calibration establishes its indication relative to a known reference.
A meter has ±1% of full-scale accuracy on a 100 V range. What is its maximum stated absolute error?
A ±1 V
B ±0.01 V
C ±10 V
D ±100 V
Answer: A. ±1 V. One percent of the 100 V full-scale value is 1 V.
That same meter reads 20 V with ±1 V maximum error. What percentage of the reading is ±1 V?
A 1%
B 0.2%
C 20%
D 5%
Answer: D. 5%. 1 V divided by 20 V is 0.05, or 5%.
Why can an average-responding AC meter misread a distorted waveform?
A Its sine calibration
B Its high bandwidth
C Its four-wire leads
D Its digital display
Answer: A. Its sine calibration. A sine-calibrated average reading does not generally equal the true RMS of a distorted wave.
Which instrument specification matters when measuring nonsinusoidal AC heating effect?
A Mirror scale
B True RMS capability
C Zero-center pointer
D Revolution constant
Answer: B. True RMS capability. True RMS measures the effective value associated with heating.
A 10× oscilloscope probe usually reduces circuit loading by doing what?
A Shorting the input
B Removing the ground
C Increasing input impedance
D Halving frequency
Answer: C. Increasing input impedance. A compensated 10× probe presents higher effective input resistance and lower capacitance.
On a scope set to 2 V/div, a trace spans 3 vertical divisions. What is its displayed peak-to-peak voltage?
A 1.5 V
B 6 V
C 3 V
D 12 V
Answer: B. 6 V. Three divisions times 2 V/div gives 6 V peak to peak.
A waveform repeats every 5 ms. What is its frequency?
A 5 Hz
B 200 Hz
C 50 Hz
D 500 Hz
Answer: B. 200 Hz. Frequency is 1/0.005 s, or 200 Hz.
What prevents aliasing before an ADC samples a high-frequency signal?
A Anti-alias filter
B Shunt resistor
C Mirror scale
D CT burden
Answer: A. Anti-alias filter. A low-pass anti-alias filter limits frequencies above the usable Nyquist band.
For a 50 Hz signal, which sampling rate exceeds its 100 Hz Nyquist minimum?
A 500 samples/s
B 50 samples/s
C 80 samples/s
D 100 samples/min
Answer: A. 500 samples/s. 500 samples/s is greater than twice 50 Hz.